Bedside Topline
- What It Does
- A helium balloon in the descending aorta inflates in diastole to raise coronary perfusion and deflates just before systole to lower afterload.
- Key Settings & Targets
- Trigger: ECG or arterial pressure
- Ratio: 1:1 to start, then step down
- Inflate at the dicrotic notch; deflate before systole
- MAP target ≥65 mmHg
- Check Every Hour
- Distal pulses and perfusion in both legs, left radial pulse and neuro check, insertion site, console timing and augmentation, helium level, MAP and urine output.
- Top Alarm → First Move
- Loss of trigger or pressure: check leads or transducer, then switch to the arterial trigger if the trace is clear.
- Never
- Run the pump with blood in the gas line
- Leave the balloon static in blood
- Reduce fill volume to wean
- Clamp or pull the catheter without an order
- Escalate Now If
- Blood in the helium line or gas loss
- Cool, pale or painful leg
- New neurologic deficit
- Late deflation not fixed by timing
Overview & How It Works
An intra-aortic balloon pump (IABP) is temporary mechanical support. A helium-filled balloon on a catheter sits in the descending thoracic aorta. It inflates in diastole and deflates just before systole. Inflation raises diastolic pressure and coronary perfusion. Deflation lowers the afterload the left ventricle has to eject against.
Key Components
| Component | Where It Sits | What It Does |
|---|---|---|
| Balloon catheter | Descending thoracic aorta; tip about 1–2 cm distal to the left subclavian origin (per manufacturer IFU; confirm on chest X-ray) | Displaces aortic blood during inflation and empties during deflation |
| Sheath and femoral access | Common femoral artery (usual route) | Introduces the catheter; the leg stays straight while in place |
| Console | Bedside cart or stand | Times inflation and deflation, shows ECG, arterial and balloon traces, and sets the assist ratio |
| Helium supply | Cylinder on the console | Fills and empties the balloon quickly; helium is inert and diffuses out rapidly if it leaks |
| Pressure lumen | Central lumen of the catheter and its tubing | Measures arterial pressure at the balloon for the console display and timing |
| Trigger input | ECG leads or arterial pressure line | Gives the console the cardiac cycle signal it times each beat from |
Flow Path
- The console watches for a trigger: an ECG R-wave, an arterial upstroke or a pacer signal.
- Helium fills the balloon at the dicrotic notch. That is when the aortic valve closes.
- The balloon stays full through diastole, pushing blood toward the coronary arteries.
- The balloon empties just before the next systolic upstroke. The left ventricle ejects against lower aortic pressure.
Physiology & Hemodynamic Effects
What It Does
| Parameter | Effect | Why It Matters |
|---|---|---|
| Diastolic aortic pressure | ↑ | Raises coronary perfusion pressure during diastole, when coronary flow occurs |
| LV afterload | ↓ | Less pressure to eject against; lower wall stress |
| Isovolumetric contraction | ↓ | Shorter time at high LV pressure lowers myocardial O₂ demand |
| Stroke volume and cardiac output | ↑ (usually modest) | Benefit depends on rhythm, native LV function and timing |
| Left-sided filling pressures | ↓ (variable) | Offloading the LV can lower wedge and LA pressures in some patients |
| Systemic vascular resistance | ↓ (functional) | Afterload reduction lowers SVR, so MAP can rise only if flow also improves |
What It Depends On
- Rhythm: you need an organized rhythm with a trigger you can use. Ventricular fibrillation or pulseless ventricular tachycardia gives little or no benefit.
- Native LV function: the biggest effect shows up when the ventricle still ejects.
- Timing: inflate at the dicrotic notch and deflate before systole. Timing errors reduce the benefit or reverse it.
- Balloon volume and position: a poorly filled or misplaced balloon augments weakly.
- Aortic compliance and pressure: stiff aortas and hypertension blunt the augmented diastolic pressure.
Why Helium
Helium is light, so the balloon fills and empties fast. It is also inert, but it is absorbed fast once it reaches the circulation. A rupture with gas loss, or blood in the gas line, is an emergency.
Types, Modes & Configurations
| Type / Mode | Route / Configuration | Support or Output | Best For | Key Limitation / Risk |
|---|---|---|---|---|
| Femoral IABP (standard) | Femoral artery; balloon in the descending thoracic aorta | Modest augmentation; modest rise in cardiac output | Short-term bridge in selected cardiogenic shock with an organized rhythm, or a bridge for a mechanical complication of MI | Limb ischemia, bleeding, thrombosis, balloon rupture; no routine mortality benefit in shock |
| Asynchronous (internal) IABP | Same catheter; console fires on a fixed internal rate | No true counterpulsation; keeps the balloon moving | Only when there is no usable ECG or pressure trigger, such as during arrest | Timing is not linked to the heartbeat, so it can worsen afterload |
Trigger Modes
- ECG (R-wave): the usual trigger, as long as the QRS is tall and consistent.
- Arterial pressure: reads the arterial upstroke. It helps with atrial fibrillation, ectopy or a noisy ECG.
- Pacer: for paced rhythms. Pick ventricular or atrial pacer input per the console setting.
- Internal (asynchronous): the last resort, for when neither ECG nor pressure gives a reliable trigger. It is not true counterpulsation.
Assist Ratios
- 1:1: the balloon inflates every beat. Start here for shock and transport.
- 1:2: every second beat is assisted. You use it to compare assisted and unassisted beats, and early in weaning.
- 1:3: minimal support. A late weaning step in stable patients.
Indications, Contraindications & Evidence
Indications
- Cardiogenic shock complicating acute MI: it may bridge you while the cause is treated. Current guidelines do not recommend routine use (see Key Evidence).
- Mechanical complication of MI: acute severe mitral regurgitation or ventricular septal rupture. It is a bridge to surgery or definitive therapy.
- Refractory ischemia: ischemia that continues despite medical therapy while you wait for revascularization. Selected patients only.
- Post-cardiotomy low output: trouble weaning off cardiopulmonary bypass, or low output after cardiac surgery. Practice varies by center.
- Bridge: to another support device (LVAD, ECMO or a microaxial pump) or to surgery. The team decides.
Contraindications
| Absolute | Relative |
|---|---|
| Moderate to severe aortic regurgitation | Uncontrolled bleeding or severe coagulopathy (insertion and removal risk) |
| Aortic dissection | Severe peripheral arterial or iliac occlusive disease that prevents safe insertion |
| Thoracic or abdominal aortic aneurysm that the balloon cannot safely traverse or sit in | Thrombocytopenia or heparin-induced thrombocytopenia |
Sepsis or uncontrolled bacteremia (device infection risk) | |
Sustained tachyarrhythmia or a rhythm that prevents reliable triggering | |
Irreversible end-stage disease with no recovery, transplant or LVAD option |
Limitations
- Needs an organized ventricular rhythm. Little benefit in ventricular fibrillation or pulseless ventricular tachycardia.
- Does not directly support the right ventricle.
- Augmentation stays modest next to ECMO or microaxial pumps.
- Benefit depends on timing. Rhythm changes and ectopy reduce effectiveness.
- In IABP-SHOCK II, routine use in cardiogenic shock did not improve survival.
Key Evidence
| Trial / Guideline (Year) | Population | Bottom Line |
|---|---|---|
| IABP-SHOCK II (2012; 12-month 2013; 6-year 2019) | Randomized patients with acute MI and cardiogenic shock; IABP with standard care versus standard care alone | No reduction in 30-day mortality; no difference at 12 months or 6 years |
| BCIS-1 (2010; long-term follow-up 2013) | Randomized patients with ejection fraction under 30% and severe coronary disease undergoing high-risk PCI; elective IABP versus no planned IABP | No difference in major adverse cardiac and cerebrovascular events (MACCE) at hospital discharge. At a median of 51 months, all-cause mortality was lower with elective IABP (HR 0.66, 95% CI 0.44–0.98), a long-term follow-up finding |
| DanGer Shock (2024) | Patients with STEMI-related cardiogenic shock; microaxial flow pump (Impella CP) plus standard care versus standard care alone | Microaxial pump plus standard care in the selected trial population lowered 180-day all-cause mortality (45.8% vs 58.5%; HR 0.74) but caused more serious adverse events (composite 24.0% vs 6.2%: severe bleeding, limb ischemia, hemolysis, device failure or worsening aortic regurgitation). See the Impella page |
| 2023 ESC ACS guideline | Acute coronary syndrome, including cardiogenic shock | States that routine IABP is not recommended for cardiogenic shock complicating AMI, unless a mechanical complication is present |
| 2025 ACC/AHA ACS guideline | Acute coronary syndrome, including cardiogenic shock | Routine IABP in AMI-related cardiogenic shock is not recommended (Class 3, no benefit; downgraded from Class 2a in 2013). The same guideline gives microaxial flow pumps Class 2a (reasonable) for selected STEMI patients with severe or refractory shock. Class wording as summarized in the JACC companion editorial |
The SCAI SHOCK classification stages shock severity, from stage A to stage E. The stage tells you when and how to escalate support.
Setup, Settings & Targets
Targets
| Parameter | Typical Target | If Below | If Above |
|---|---|---|---|
| MAP | ≥65 mmHg | Check augmentation, fill and rhythm; team decides on vasoactive support | Usually acceptable; check afterload and vasoactive dose |
| Augmented diastolic pressure | Taller than the unassisted systolic peak | Check timing, fill and balloon position | Not usually a problem; recheck if waveform looks wrong |
| Assisted systolic pressure | Slightly lower than unassisted systolic | Expected finding with good afterload reduction | Equal to or above unassisted: suspect early deflation (no afterload reduction) |
| Assisted end-diastolic pressure | Lower than unassisted end-diastolic | Expected finding; a sharp U-shaped dip after the augmented peak suggests early deflation | Equal to or above unassisted: suspect late deflation |
Initial Settings
- Trigger: ECG when the R-wave is clean and tall. Arterial pressure for atrial fibrillation, ectopy or pacing.
- Ratio: start at 1:1. Step down only after the patient is stable.
- Timing: inflate at the dicrotic notch. Deflate before the next systolic upstroke. Adjust with the console timing controls.
- Fill and helium: check fill status and gas level on the console at each check.
Sizing & Positioning
- Balloon volume: matched to patient height on the manufacturer chart (per manufacturer IFU).
- Tip position: about 1–2 cm distal to the left subclavian origin. Confirm it on chest X-ray (per manufacturer IFU).
Anticoagulation
The team sets anticoagulation targets. You will find them in the Anticoagulation & Hematology section.
Calculations & Derived Values
| Variable | Formula | Why You Care |
|---|---|---|
| MAP | DBP + ⅓ (SBP − DBP) | Perfusion target (≥65 mmHg) and the pressure the balloon must support |
| Cardiac power output | MAP × CO ÷ 451 | Combines pressure and flow; a low value signals shock severity and prompts an escalation discussion |
| Cardiac index | CO ÷ BSA (L/min/m²) | Used as a readiness threshold for weaning (protocol-dependent) |
Waveforms & Console Interpretation
Normal
- Inflation: a sharp V at the dicrotic notch. The augmented diastolic peak follows right after it.
- Augmented diastolic peak: taller than the unassisted systolic peak.
- Assisted end-diastolic pressure: lower than the unassisted end-diastolic pressure. It is measured just before the next upstroke.
- Assisted systolic pressure: slightly lower than the unassisted systolic peak.
Abnormal Patterns
| Pattern | What You See | Cause | Fix |
|---|---|---|---|
| Early inflation | Balloon inflates before the dicrotic notch; augmented peak merges with systolic | Inflation timed before aortic valve closure | Delay inflation to the dicrotic notch |
| Late inflation | Dicrotic notch visible before the augmented peak; small augmented peak | Inflation timed after the notch | Advance inflation so it starts at the notch |
| Early deflation | Assisted systolic pressure equal to or above unassisted (no afterload reduction); sharp drop after the augmented peak; assisted end-diastolic pressure equal to or below unassisted (U-shaped dip) | Balloon empties before end-diastole; can cause retrograde coronary and carotid flow | Delay deflation so it ends at end-diastole, just before the systolic upstroke |
| Late deflation (among the most harmful) | Assisted end-diastolic pressure equal to or above unassisted; widened augmented diastolic peak; slowed assisted systolic upstroke as the balloon is still deflating into systole | Balloon is still inflated after the aortic valve opens, raising afterload | Advance deflation so the balloon is empty by end-diastole, before ejection |
Monitoring & Ongoing Care
| Frequency | What to Check | Expected / Action |
|---|---|---|
| Hourly | Distal pulses, color, temperature, capillary refill, sensation and movement in both legs | Symmetric perfusion; any new change needs immediate assessment and provider notice |
| Hourly | Left radial pulse, left arm BP and color; neuro check (speech, facial symmetry, limb strength) | Symmetric arm perfusion and neuro exam; a weak or absent left radial pulse, lower left arm BP or a new deficit suggests upward migration: notify the provider now |
| Hourly | Augmented peak, timing, trigger, fill and helium level on the console | Augmentation present on every assisted beat; loss of augmentation needs troubleshooting |
| Hourly | MAP, heart rate and rhythm | MAP ≥65 mmHg; new rhythm changes need a timing review |
| Hourly | Insertion site for bleeding or hematoma; urine output | Stable site; urine output per unit target; falling output needs review |
| Per unit protocol | CBC and platelets; aPTT or ACT per heparin protocol | Per order and protocol. A platelet fall of 50% or more from baseline during heparin exposure: notify the team for heparin-induced thrombocytopenia evaluation |
| Per unit protocol | Creatinine, plasma free hemoglobin and LDH when hemolysis is suspected; chest X-ray if position is in question | Rising values or a moved catheter need provider review |
Positioning & Mobility
- Keep the leg with the femoral catheter straight and limit hip flexion per unit protocol.
- Keep head-of-bed elevation at or below 30 degrees for femoral access (higher angles can strain or kink the catheter); log-roll rather than sit the patient up when possible.
- Expect no ambulation with a femoral catheter. Passive range of motion on the unaffected leg, per team.
Anticoagulation & Hematology
| Strategy / Agent | Typical Target | Monitoring | Watch For |
|---|---|---|---|
| Unfractionated heparin High-alert | Set by team protocol (no universal target) | ACT or aPTT per protocol | Bleeding; balloon thrombosis when flow is poor; heparin-induced thrombocytopenia |
| Bivalirudin or argatroban High-alert | Per protocol | aPTT per protocol | Used when heparin cannot be given, such as heparin-induced thrombocytopenia |
Hemolysis
- Signs: dark or tea-colored urine, rising LDH, falling haptoglobin, rising plasma free hemoglobin, rising bilirubin.
Bleeding & Platelets
- Falling platelet count with heparin exposure: consider heparin-induced thrombocytopenia and notify the team before further heparin.
- Protamine reverses heparin only under provider order. It can cause severe hypotension, cardiovascular collapse, pulmonary vasoconstriction or anaphylactoid reactions, especially with rapid infusion.
Key Medications & Drug Considerations
| Medication | Role Here | Typical Adult Dose / Target | Device-Specific Notes |
|---|---|---|---|
| Heparin High-alert | Prevents thrombus on the catheter | Per protocol; ACT or aPTT target set by team (no fixed IABP target) | Dose and monitoring follow the unit protocol; bleeding and thrombocytopenia checks apply |
| Norepinephrine High-alert | Vasopressor when the pump alone does not hold MAP | Typical 0.05–0.5 mcg/kg/min IV infusion; titrate to MAP target (≥65 mmHg) per protocol | Excess vasoconstriction raises afterload and works against the pump |
| Dobutamine High-alert | Inotrope for low output | 2–20 mcg/kg/min IV infusion (usual; maximum per protocol) | Can increase heart rate and ectopy, which makes timing harder |
| Milrinone High-alert | Inodilator for low output | Label: 50 mcg/kg IV over 10 min, then 0.375–0.75 mcg/kg/min IV infusion | Vasodilation can lower MAP; renally cleared, so reduce in renal impairment per label |
| Amiodarone High-alert | Antiarrhythmic for VT or VF | Pulseless VF/VT: initial 300 mg IV/IO push, then repeat 150 mg once (ACLS); stable VT: 150 mg IV over 10 min | Treat the rhythm first; a balloon does not replace ACLS care |
Doses vary by protocol; follow local protocols and medical direction. Pediatric dosing is in Special Populations.
Alarms & Troubleshooting
| Alarm / Problem | What You See | Likely Causes | First Actions |
|---|---|---|---|
| Blood in helium line | Red or pink fluid in the gas tubing | Balloon rupture | 1) Stop the pump per protocol. 2) Notify the provider immediately. 3) Do not restart without an order. |
| Rapid helium loss | Falling gas level; repeated fill alarms | Balloon rupture or a leak at a connection | 1) Check tubing and connections for kinks or disconnections. 2) If loss continues, treat as balloon rupture. |
| Loss of trigger | No trigger; erratic or missing inflation | Lead off, poor R-wave, lead or cable fault | 1) Check leads and cables. 2) Switch to arterial trigger if the arterial trace is clear. |
| Loss of pressure waveform | Flat or absent arterial or balloon trace | Kinked or clotted line, air in tubing, transducer level or zero | 1) Check the pressure bag, tubing and stopcocks. 2) Re-zero and level the transducer. 3) Use an alternate arterial line if available. |
| Poor augmentation | Low augmented peak; no clear V-notch | Wrong timing, low fill, rhythm change, catheter migration | 1) Check trigger and timing. 2) Check fill and helium. 3) Review rhythm and tell the team. |
| Timing error | Early or late inflation or deflation pattern | Timing settings out of step after a rhythm or heart rate change | Adjust timing per the waveform table; recheck the next several beats. |
- Never run the pump with blood in the gas line.
- Never leave the balloon static unless the team has ordered it stopped.
- Never reduce balloon fill volume to wean.
- Never move, clamp or disconnect the catheter without a provider order.
Complications & Red Flags
| Complication | How You Recognize It | Immediate Action | Prevention |
|---|---|---|---|
| Aortic injury or dissection (insertion) | Sudden back or chest pain; pulse or BP difference | Stop the procedure; notify the team immediately | Gentle advancement; check the aortic anatomy before insertion |
| Bleeding or hematoma (insertion and during use) | Oozing or swelling at the sheath site; falling hemoglobin | Apply pressure; notify the provider | Site checks per protocol; anticoagulation per protocol |
| Limb ischemia (during use) | Cool, pale, painful leg; loss of distal pulses | Notify the provider now; check the leg straight and pulses | Keep the leg straight; hourly distal checks |
| Balloon rupture and gas embolism (during use) | Blood in the helium line; gas loss; new neurologic change | Stop the pump; notify the provider immediately | Check tubing and connections; avoid kinks during transfers |
| Thrombosis or embolism (during use) | New neurologic deficit; cold limb; loss of augmentation from a clotted balloon | Notify the provider; assess neurologic status | Anticoagulation per protocol; keep the balloon moving |
| Hemolysis and thrombocytopenia (during use) | Dark urine; rising LDH or plasma free hemoglobin; falling platelets | Notify the team; labs per order | Lab monitoring; check position |
| Downward migration: renal or mesenteric compromise (during use) | Falling urine output; rising creatinine; new abdominal pain or distension | Notify the team now; position check by imaging | Confirm tip position on X-ray after insertion and after any suspected movement |
| Upward migration: left arm ischemia or stroke (during use) | Weak or absent left radial pulse; lower left arm BP; cool or pale left arm; new neurologic deficit | Notify the provider now; position check by imaging; treat a neurologic change per stroke protocol | Confirm tip position on X-ray; hourly left arm pulse, BP and neuro checks |
| Infection (during use) | Fever; redness or drainage at the site | Culture and notify the provider | Aseptic site care; remove as soon as clinically appropriate |
| Pseudoaneurysm or AV fistula (removal) | Pulsatile mass; bruit; pain at the site | Notify the provider; site evaluation | Hemostasis per protocol after removal |
| Catheter entrapment (removal) | Resistance during removal | Do not force; notify the team | Remove only under provider order |
- Blood in the helium line or unexplained gas loss.
- Weak or absent left radial pulse, or lower left arm BP.
- Cool, pale, pulseless or painful leg.
- New neurologic deficit.
- Sudden loss of augmentation or a waveform change not fixed by timing adjustment.
- Falling MAP or urine output despite the pump.
Emergencies: Arrest, Power Loss & Device Failure
Cardiac Arrest or VF on the IABP
- Start CPR and defibrillate per the AHA 2025 adult ACLS algorithm. Do not delay life-saving care for the pump.
- Defibrillate per ACLS; follow your console's IFU for the pump during defibrillation. Do not delay shocks for the pump.
- Set the trigger for CPR: use the pressure trigger only if compressions produce a usable arterial waveform; otherwise use internal (asynchronous) mode per the console instructions.
- Notify the MCS team and document the trigger and timing change.
Power or Console Failure
- Keep the patient alive first. Assess perfusion and rhythm.
- Switch to battery or the backup console if the unit has one; check the battery charge.
- Follow the manufacturer IFU and the MCS team for manual or backup operation. A balloon left static in blood risks thrombus: restore pumping quickly or escalate.
Dislodgement or Catheter Migration
- Stop the pump and check the site and distal perfusion.
- Do not advance, withdraw or clamp the catheter without an order.
- Call the MCS team for imaging and repositioning.
Balloon Rupture or Air in the Circuit
- Stop the pump per the device instructions.
- Watch for gas embolism signs: new neurologic change, chest pain or hypotension.
- Notify the provider immediately and prepare for transfer to the MCS team.
Weaning & Discontinuation
Readiness Criteria
- MAP ≥65 mmHg on low or no vasopressor support.
- Cardiac index about 2.2 L/min/m² or higher (protocol-dependent).
- Lactate normalizing; perfusion improving.
- Improving LV function on echo; cause of shock treated.
Weaning Steps
- Confirm readiness with the team.
- Reduce the ratio stepwise: 1:1 to 1:2 to 1:3, keeping the balloon fill unchanged.
- Reassess MAP, perfusion, lactate and urine output at each step.
- If MAP, perfusion or symptoms worsen, step back to the prior ratio.
- Remove only after a stable period on the lowest support level, per protocol.
Failure Criteria: Step Back Up
- Falling MAP or new vasopressor need.
- Rising lactate or falling urine output.
- New arrhythmia or loss of augmentation.
Removal & Post-Removal Care
- The team removes it, under order. Heparin is held and coagulation is checked per protocol before removal.
- After removal: manual pressure at the site for the protocol-defined time, then hemostasis, distal pulse and limb checks, and bed rest per protocol.
- A balloon must not sit static in blood past the manufacturer time limit. If the pump cannot be restarted, escalate to the team.
Transport Considerations
Before You Move
- Check battery charge and confirm backup power is available.
- Check the helium cylinder supply and how long it will last.
- Secure the catheter, tubing and cables.
- Pack spare tubing, a spare pressure bag and the backup console or manual protocol.
- Confirm trained personnel are with the patient and the receiving team has been briefed.
If You Are Not the Device Specialist
- Watch: distal perfusion of the catheter leg, the arterial trace, and the console alarm screen.
- Never touch: trigger, ratio or timing settings unless trained; never clamp or pull the catheter.
- Call: the MCS team or provider for any waveform change, blood in the gas line or limb change.
- CPR and defibrillation: follow ACLS; see Emergencies for the device-specific steps.
During Transport
- Keep the catheter leg straight and the tubing secured through every transfer.
- Recheck distal perfusion after each move.
Air Medical & Altitude
Helium expands as ambient pressure falls (Boyle's law). So after every altitude change, check balloon fill and console status. Confirm with the flight team per the device IFU.
Handoff
- Mode, trigger, ratio, timing and fill settings.
- Latest labs, anticoagulation and platelet trend.
- Recent alarms or complications.
- Access sites, distal limb checks and the last time they were done.
- Goals of care and the device plan.
Special Populations
Pediatric & Neonatal
Only specialized centers place balloon pumps in children. Catheter size follows the child's vessels, set by manufacturer and center protocol. For pediatric shock support and vascular access, see pediatric shock and vascular access. Weight-based dosing governs the pediatric doses of the medications above. Look them up on the medication pages.
Other
- Older adults and peripheral arterial disease: higher risk of access-site and limb complications. Check pulses before insertion.
- Pregnancy: data are limited. The decision rests with the team.
Bedside Pearls & Teaching Script
Systematic Walk-Through
- Indication: why the balloon is in place and whether it is a bridge.
- Mode: trigger, ratio and whether augmentation is visible.
- Settings: inflation and deflation timing, fill and helium level.
- Anticoagulation: target and latest platelets and aPTT.
- Complications: limb, site, hemolysis and neurologic check.
- Today's goal: weaning, escalation or transfer.
Pearls
- A taller augmented peak and a lower assisted end-diastolic pressure: that is what good timing looks like.
- Of the timing errors, late deflation and early inflation are among the most harmful.
- A static balloon is a thrombus risk. Keep the pump running unless it is stopped on order.
- Severe aortic regurgitation: the contraindication most often missed.
- Routine IABP in cardiogenic shock has not improved survival. The goal is a bridge.
References
- Thiele H, et al. (2012). Intraaortic balloon support for myocardial infarction with cardiogenic shock. New England Journal of Medicine, 367, 1287–1296. https://doi.org/10.1056/NEJMoa1208410
- Thiele H, et al. (2013). Intra-aortic balloon counterpulsation in acute myocardial infarction complicated by cardiogenic shock (IABP-SHOCK II): final 12 month results of a randomised, open-label trial. Lancet, 382, 1638–1645. https://doi.org/10.1016/S0140-6736(13)61783-3
- Perera D, et al. (2013). Long-term mortality data from the balloon pump-assisted coronary intervention study (BCIS-1): a randomized, controlled trial of elective balloon counterpulsation during high-risk percutaneous coronary intervention. Circulation, 127(2), 207–212. https://doi.org/10.1161/CIRCULATIONAHA.112.132209
- Thiele H, et al. (2019). Intraaortic balloon pump in cardiogenic shock complicating acute myocardial infarction: long-term 6-year outcome of the randomized IABP-SHOCK II trial. Circulation, 139, 395–403. https://doi.org/10.1161/CIRCULATIONAHA.118.038201
- Perera D, et al. (2010). Elective intra-aortic balloon counterpulsation during high-risk percutaneous coronary intervention: a randomized controlled trial. JAMA, 304(8), 867–874. https://doi.org/10.1001/jama.2010.1190
- Møller JE, et al. (2024). Microaxial flow pump or standard care in infarct-related cardiogenic shock (DanGer Shock). New England Journal of Medicine, 390(15), 1382–1393. https://doi.org/10.1056/NEJMoa2312572
- Byrne RA, et al. (2023). 2023 ESC Guidelines for the management of acute coronary syndromes. European Heart Journal: Acute Cardiovascular Care, 13(1), 55–161. https://doi.org/10.1093/ehjacc/zuad107
- Rao SV, et al. (2025). 2025 ACC/AHA/ACEP/NAEMSP/SCAI guideline for the management of patients with acute coronary syndromes. Journal of the American College of Cardiology, 85, 2135–2237. https://doi.org/10.1016/j.jacc.2024.11.009
- Murugiah K, McDonagh TA, Cohen DJ, Dhruva SS. (2025). Mechanical circulatory support in acute myocardial infarction-cardiogenic shock: 2025 acute coronary syndrome guideline in context. Journal of the American College of Cardiology, 85(22), 2103–2106. https://doi.org/10.1016/j.jacc.2025.04.025
- Naidu SS, et al. (2022). SCAI SHOCK stage classification expert consensus update: a review and incorporation of validation studies. Journal of the American College of Cardiology, 79(9), 933–946. https://doi.org/10.1016/j.jacc.2022.01.018
- Motia N, Marko V, Karlsen MW. (2024). Complications associated with intra-aortic balloon pump treatment in critically ill patients: a systematic review. Nursing in Critical Care, 29(6), 1768–1780. https://doi.org/10.1111/nicc.13163
- American Heart Association (2025). Part 1: Executive Summary: 2025 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. https://doi.org/10.1161/cir.0000000000001372
Medical Disclaimer
- For Educational Purposes Only: This content is intended for educational reference and should not be used for clinical decision-making.
- Not a Substitute for Professional Judgment: Always consult your local protocols, institutional guidelines, and supervising physicians.
- Verify Before Acting: Users are responsible for verifying information through authoritative sources before any clinical application.