Bedside Topline

What It Does
A temporary pulse generator paces the heart through a transvenous catheter in the right ventricle or through epicardial wires placed at surgery.
Key Settings & Targets
  • Rate: 60–80/min (typical; follow the order)
  • Output: above threshold by the protocol margin
  • Capture: ECG and pulse confirmed
Check Every Hour
Capture on the ECG, pulse with each paced beat, rate and mode, battery, site and connections.
Top Alarm → First Move
Failure to capture: check the pulse, raise output in steps, check the connections, then call the team.
Never
Turn the generator off in a pacer-dependent patient; push or pull a lead; change the mode or output without an order.
Escalate Now If
Loss of capture with a falling blood pressure, pulseless electrical activity, muffled heart sounds, or a dislodged lead.

Overview & How It Works

An external pulse generator sends an electrical impulse to the heart. The route is a catheter placed in the right ventricle (transvenous) or wires sewn to the heart surface (epicardial). Use it when the heart cannot generate a rate or conduction that keeps the patient perfused. It bridges the time until a permanent solution or recovery. It is a bridge, not a fix.

Key Components

ComponentWhere It SitsWhat It Does
External pulse generator
Bedside, on a stand or in a transport case
Sets rate and output, senses the native rhythm, and shows the battery state
Transvenous pacing catheter
Right atrium and right ventricle, through an entry vein
Delivers the impulse at the right ventricular endocardium
Epicardial wires
Sewn to the atrial or ventricular epicardium at surgery, brought out through the chest wall
Deliver the impulse after cardiac surgery
Cables and connectors
Between the leads and the generator
Carry the signal; loose or wet connections cause sensing and capture faults
Entry site and dressing
Neck, chest, groin or the surgical wire exit site
Infection and dislodgement risk; check every shift

Flow Path

  1. Generator: sets the rate, output and sensitivity.
  2. Lead: the impulse runs down the wire or catheter.
  3. Myocardium: once the impulse clears the threshold, the chamber depolarizes. That is capture.
  4. Sensing: it listens for the native beat, then inhibits or triggers as programmed.
Key Point: The spike is not proof of pacing. The proof is a wide paced complex on the ECG after the spike. When the patient has a pulse, check that pulse too.

Transcutaneous pacing runs on skin pads with its own settings. The transcutaneous pacing page covers that route.

Physiology & Hemodynamic Effects

What It Does

ParameterEffectWhy It Matters
Heart rate
↑ to the set rate when the native rate is slower
Restores a rate that supports cardiac output in bradycardia or heart block
AV synchrony
↑ with atrial-tracking modes (AAI, DDD) when the atria are intact
Atrial contraction adds to ventricular filling; ventricular-only pacing (VVI) loses it
Cardiac output
↔ or ↑ when pacing restores a perfusing rate; ↓ with pacing-induced dyssynchrony
Output depends on rate, rhythm and filling, not on the spike alone
Ventricular activation
Altered (wide paced QRS) when pacing from the RV apex
Dyssynchrony can lower function in some patients, especially with reduced EF
Pacing threshold
↑ early after insertion and with ischemia, acidosis, hyperkalemia or lead displacement
A rising threshold means the output may no longer capture

What It Depends On

  • Lead position and tissue contact. The threshold and sensing change when the lead moves.
  • Electrolytes and acid-base status (potassium, pH, calcium)
  • Native conduction. A pacer-dependent patient has no reliable escape rhythm to fall back on.
  • Drugs that change the threshold or conduction (antiarrhythmics, some sedatives)
  • Pacing mode and whether the atria are paced or only sensed
Clinical Pearl: Wrong mode, wrong result. A patient paced in VVI while in sinus rhythm with good AV conduction can turn hypotensive once pacing starts. Look at the rhythm and the pulse. The rate number is only part of the picture.

Types, Modes & Configurations

Type / ModeRoute or ConfigurationBest ForKey Limitation / Risk
Transvenous (RV)
Internal jugular, subclavian or femoral vein; tip in the RV apex
Bradycardia or heart block where the transcutaneous route fails or is not tolerated
Dislodgement, perforation, pneumothorax, infection, arrhythmia during placement
Epicardial (surgical)
Atrial and ventricular wires sewn at surgery; exit through the chest wall
Bradyarrhythmia after cardiac surgery, with the wires left in place
Wires are unipolar or bipolar by design; wrong connections cause failure to capture; removal can bleed
Transcutaneous (bridge)
Pads on the chest and back
Immediate bridge while a transvenous or epicardial route is arranged
Painful, high output, capture may be unreliable

Pacing Mode Letters (NBG Code)

LetterPositionMeaning
1st
Chamber paced
V = ventricle, A = atrium, D = both
2nd
Chamber sensed
V, A or D; the generator looks for the native beat here
3rd
Response to a sensed beat
I = inhibit (wait), T = trigger (pace now), D = both

Common Modes

  • VVI: ventricular pacing. A native ventricular beat inhibits it. This is the usual temporary mode.
  • AAI: atrial pacing with intact AV conduction.
  • DDD: dual-chamber. It tracks the atria while AV conduction is intact. It needs both leads.
  • Asynchronous (VOO or DOO): it paces at a fixed rate and ignores the native rhythm. It is used to stop oversensing or inhibition during electrocautery. The trap: a spike that lands on a native T wave can cause R-on-T and VF.
Electrical Safety (Epicardial and Transvenous Leads): Exposed lead terminals and wire ends can conduct current to the heart. That is microshock. Handle connectors and exposed terminals with insulated gloves. Cap or insulate unused or disconnected wire ends, for example with a finger cot or dry dressing. Keep the generator and cables dry. Keep other mains-powered equipment from touching the leads.
Warning: With a native rhythm present, asynchronous pacing can fire on the T wave. Use it only with a physician order and on a monitor. Turn it back to demand mode when the procedure ends.

Indications, Contraindications & Evidence

Indications

  • Symptomatic bradycardia with poor perfusion that does not respond to drug therapy, per the local bradycardia algorithm
  • Symptomatic Mobitz II or complete (third-degree) AV block, especially with a wide QRS or an unreliable escape rhythm
  • Sinus node dysfunction with symptoms, as a bridge to a permanent device
  • Bradyarrhythmia after cardiac surgery, with epicardial wires in place
  • Bradycardia from drug toxicity, when drug therapy has failed and the team is arranging a pacing route

Contraindications

AbsoluteRelative (Protocol-Dependent)
No absolute contraindication in an emergency; the decision is to pace or not, based on the indication
Coagulopathy or thrombocytopenia (venous hematoma and bleeding)
Not applicable as a rule; some protocols list active line infection or bacteremia at the entry site
Tricuspid prosthetic valve, or other anatomy that makes placement difficult
Pneumothorax risk at the subclavian or internal jugular site; lead placed by an experienced operator

Limitations

  • Transvenous pacing needs an operator who can place the lead and position it by fluoroscopy or ECG guidance
  • Epicardial wires depend on where the surgeon placed them. Their function changes as the myocardium heals
  • Temporary pacing does not treat the cause of the bradycardia

Key Evidence

Guideline / Source (Year)PopulationBottom Line
2018 ACC/AHA/HRS bradycardia guideline (published 2019)
Adults with bradycardia and conduction delay
Covers the evaluation and management of symptomatic bradycardia and AV block, including temporary pacing, with its own recommendation classes
2012 ACCF/AHA/HRS focused update of the 2008 device guideline
Adults needing cardiac rhythm devices
Updates device indications and pacing practice; read the recommendation tables before applying a class
2025 AHA CPR and ECC guidelines, Part 9: Adult Advanced Life Support (2025)
Adults in cardiac arrest and peri-arrest conditions
Current AHA adult advanced life support guidance; confirm the bradycardia algorithm and drug doses in the full text before use
Chinese Medical Association Emergency Branch consensus on bedside temporary cardiac pacing (2023)
Emergency bedside temporary pacing
Covers route choice, venous access, lead selection, parameter settings, complications and post-procedure care (published in Chinese)

Protocol-dependent. The margin for pacing output and the threshold to start pacing vary between institutions. This page sets no single number for either.

Setup, Settings & Targets

Targets

ParameterTypical TargetIf BelowIf Above
Rate
60–80/min (typical; follow the order)
Check capture and sensing; call the team
Check for pacing-induced tachycardia or oversensing
Output (mA)
Set above the capture threshold by the protocol margin (protocol-dependent)
Failure to capture: raise output in steps and recheck the threshold
Output above what is needed increases the risk of tissue injury and generator drain
Sensitivity
Set so the generator sees the native beat and inhibits when it should
Undersensing: pacing competes with native beats
Oversensing: pacing is inhibited by noise or T waves
Pulse with each paced beat
Present when capture is confirmed
No pulse with spikes: pulseless electrical activity pathway
—

Initial Settings

  1. Confirm the indication and the order; set the mode the order specifies.
  2. Set the rate; start at the ordered rate.
  3. Increase output in steps until capture is seen on the ECG and the pulse is checked.
  4. Decrease output in steps to find the threshold (the lowest output with capture); set the working output above the threshold by the protocol margin.
  5. Set sensitivity so the generator senses the native rhythm; test it by watching for inhibition with a native beat.

Sizing & Positioning

  • Transvenous: fluoroscopy or the ECG pattern, plus a chest X-ray, confirm the tip position in the RV apex, per local protocol.
  • Epicardial: the surgeon records which wires are atrial and which are ventricular. Keep that record with the generator.
Note: One margin does not fit every route. Transcutaneous pacing has its own output margin, set per manufacturer and protocol (see the transcutaneous page). Transvenous and epicardial margins depend on protocol. Teaching sources often cite about twice the threshold for transvenous pacing. No primary source here fixes a number. Do not carry the transcutaneous figure over to a transvenous lead.

Waveforms & Console Interpretation

Normal

  • Capture looks like this: a spike, then a wide QRS, with the T wave pointing the opposite way, at the set rate.
  • Demand pacing goes quiet when the native beat is there and sensed. No spike.
  • In DDD, the atrial spike comes first. The ventricular spike follows after the programmed AV delay.

Abnormal Patterns

PatternWhat You SeeCauseFix
Failure to capture
Spike with no QRS after it
Output below threshold, lead displaced, perforation, hyperkalemia or acidosis, lead fracture
Check the pulse; raise output and recheck; check the lead and electrolytes; call the team
Failure to sense (undersensing)
Spikes fall on native beats; pacing competes
Sensitivity set too low, lead displaced
Adjust sensitivity; check the lead
Oversensing
Pauses; pacing inhibited by noise, T waves or muscle activity
Sensitivity set too high, electrical interference, loose connection
Adjust sensitivity; check the connections and cables
Spikes on the T wave
Spike falls in the vulnerable window of a native beat
Asynchronous mode or wrong settings
Turn to demand mode when the procedure ends; call the team
Battery or power alarm
Warning on the generator
Battery depletion, mains failure
Switch to backup power or a spare generator per protocol
Warning: No QRS and no pulse, even with a spike on the screen? That is arrest, not pacing. Start CPR and call for help.

Monitoring & Ongoing Care

FrequencyWhat to CheckAction Threshold
Hourly
Rhythm, capture on the ECG, pulse with paced beats, rate, generator mode, battery
Any loss of capture, pulse with spikes absent, or a rate change: check and call
Hourly
Insertion site, dressing, connections and cables, lead secured
Bleeding, drainage, loose connection, lead movement: call
Every 4–6 h
Threshold check per protocol; electrolytes (potassium, magnesium, calcium)
A rising threshold or abnormal electrolytes: treat per order and recheck
Daily
Chest X-ray or other imaging for lead position, per protocol
Position change: notify the team before any repositioning
Daily
Blood tests: platelets, aPTT or anti-Xa per order
Platelets or coagulation outside the order range: notify the team

Positioning & Mobility

  • Leads stay secured. On every move, protect them and keep the generator on the bed or in the transport case
  • The entry site decides how you position the limb or neck. Avoid pulling on the cable
Clinical Pearl: Every move, position change or sedation change earns a capture check on the ECG. So does any change in rhythm.

Anticoagulation & Hematology

Strategy / AgentTypical TargetMonitoringWatch For
Unfractionated heparin (if ordered)
Per order and protocol
aPTT or anti-Xa per order
Venous hematoma, pericardial bleeding
Platelet count
Per protocol
Daily CBC
Thrombocytopenia increases hematoma risk
Antiplatelet therapy
Per order
Per order
Bleeding at the entry site

Tag heparin High-alert. Refer to the heparin page and the aPTT page.

Bleeding & Platelets

  • Venous puncture hematoma: pressure and a dressing, check the neck or groin each shift
  • Pericardial bleeding: falling blood pressure and a rising heart rate after lead placement; call immediately
  • Thrombocytopenia: check the platelet count (see the platelets page)

Key Medications & Drug Considerations

MedicationRole HereDose / TargetNotes
Atropine
First drug for sinus and AV-nodal bradycardia before pacing, per the local bradycardia algorithm
Symptomatic bradycardia: give per the current AHA ACLS bradycardia algorithm and your local protocol, which set the dose, repeat interval and maximum. Label (coronary artery disease, section 2.4): limit the total dose to 0.03–0.04 mg/kg. Atropine is not recommended for asystole or PEA arrest.
Not a reliable step in Mobitz II or complete block; do not let it delay pacing
Dopamine
Infusion bridge while pacing is set up or when pacing fails
Bradycardia: infusion titrated to response per the current AHA ACLS bradycardia algorithm and your local protocol. The DailyMed label dose for shock (start 2–5 mcg/kg/min, maximum 50 mcg/kg/min) is for a different indication and is not a bradycardia dose.
High-alert infusion; use an infusion pump
Epinephrine infusion
Infusion bridge when dopamine is not used
Bradycardia: infusion titrated to response per the current AHA ACLS bradycardia algorithm and your local protocol
High-alert infusion
Heparin
Only if ordered for another indication
Per order
High-alert
Lidocaine
May be used for ventricular ectopy that the pacing lead provokes
Per order
Ventricular ectopy during placement is a reason to stop and reposition
Warning: No bradycardia doses live on this page. Pull the dose from the current AHA ACLS bradycardia algorithm and your local protocol. Check it against the order and pharmacy before any drug goes in. DailyMed lists other uses, bradyasystolic arrest and shock among them. Neither one is the bradycardia indication here.

Alarms & Troubleshooting

Alarm / ProblemWhat You SeeLikely CausesFirst Actions
Failure to capture
Spikes with no QRS
Output below threshold, lead displaced, electrolytes, acidosis, lead fracture or connection
1) Check the pulse 2) Raise output in steps 3) Check connections and electrolytes 4) Call the team
Failure to sense
Spikes on top of native beats
Sensitivity too low, lead position
1) Adjust sensitivity 2) Check the connections 3) Call the team
Oversensing
Pauses or inhibition with no native beat
Sensitivity too high, interference, loose cable
1) Adjust sensitivity 2) Check the cables 3) Check for interference
Loss of capture after defibrillation
No capture after a shock
Threshold change from the shock, lead injury
1) Check the pulse 2) Recheck the threshold 3) Call the team
Battery or power alarm
Generator warning
Depleted battery, power failure
1) Switch to backup or spare generator 2) Keep pacing going 3) Call the team
Hypotension after pacing starts
Blood pressure falls after the mode changes
Wrong mode for the rhythm, ventricular-only pacing with good AV conduction
1) Check the rhythm and mode 2) Call the team; mode changes need an order
Never: disconnect the generator from a pacer-dependent patient without a backup in place. Pushing or pulling a lead is off the table. Change the mode or output only with an order. Use asynchronous pacing near a native rhythm only with a physician order.
Electrical Safety: Exposed epicardial or transvenous lead terminals stay insulated and dry. Handle them with insulated gloves (see Types, Modes & Configurations).
Clinical Pearl: Patient first, then the device. Check the pulse, then the ECG capture, then the generator.

Complications & Red Flags

ComplicationHow You Recognize ItImmediate ActionPrevention
Pneumothorax (neck or chest access)
Dyspnea, falling oxygen saturation, reduced breath sounds
Call the team; oxygen; chest X-ray
Experienced operator; imaging per protocol
Lead displacement
Failure to capture or sense, a changed ECG pattern
Check the output and capture; call before repositioning
Secure the lead; monitor the position
Perforation or cardiac tamponade
Hypotension, pulseless electrical activity, muffled heart sounds, pulsus paradoxus
Call the team now; prepare for pericardiocentesis per protocol
Careful placement with imaging; watch for threshold change
Infection
Fever, redness, drainage at the site
Notify the team; cultures per order
Sterile technique; dressing care per protocol
Venous thrombosis
Arm or neck swelling, pain
Call the team
Early removal when no longer needed
Arrhythmia during placement
Ventricular ectopy or VF during placement
Stop, call, reposition per the operator
Careful advancement; monitor the ECG
Red Flags: Call Immediately falling blood pressure with pacing, pulseless electrical activity, loss of capture with an unstable patient, new muffled heart sounds, sudden dyspnea, bleeding at the entry site.

Emergencies: Arrest, Power Loss & Device Failure

Cardiac Arrest on a Temporary Pacemaker

  1. Start CPR: follow the current ACLS algorithm. The pacemaker does not replace compressions.
  2. Pads: place defibrillation pads away from the pacing lead site per the manufacturer and protocol.
  3. Shock: defibrillate per the algorithm; some protocols disconnect the leads from the generator before a shock. Follow the local protocol.
  4. After the shock: check the pulse and the capture; recheck the threshold if pacing was lost.

Loss of Capture in an Unstable Patient

  1. Check the patient: pulse, blood pressure and mental status.
  2. Check the device: connections, output, battery.
  3. Bridge: start the transcutaneous route or the infusion bridge per order.
  4. Call: the team responsible for the lead.

Generator Failure in a Pacer-Dependent Patient

  1. Do not turn the generator off.
  2. Switch to a backup or spare generator per protocol.
  3. Bridge: transcutaneous pacing or the infusion per order.
  4. Call the team now.

Dislodgement or Suspected Perforation

  1. Do not push or pull the lead.
  2. Cover the entry site with a sterile dressing.
  3. Call the team; watch for hypotension and tamponade signs.

Weaning & Discontinuation

Readiness Criteria

  • The native rhythm is stable with an adequate escape rate, per the order
  • Capture and sensing are stable on the current settings
  • The cause of the bradycardia has resolved or is controlled
  • The electrolytes are corrected

Weaning Steps

  1. Lower the rate in steps, or switch to demand mode, per the order.
  2. Watch the native rhythm and perfusion at each step.
  3. Pause pacing only under an order, with a team present and a backup ready.

Failure Criteria: Step Back Up

  • Falling blood pressure, altered mental status or chest pain with the native rate
  • Pauses or junctional or idioventricular escape that is too slow

Removal & Post-Removal Care

  • Removal per the order and the local protocol
  • The surgical team removes epicardial wires per protocol. Do not pull against resistance
  • Watch the site for bleeding. Check the chest X-ray per protocol after a transvenous removal
Protocol Varies: removal technique, timing and post-removal imaging all depend on protocol.

Transport Considerations

Before You Move

  1. Check battery charge and carry a charged spare.
  2. Confirm the mode, rate and output with the sending team; write them down.
  3. Secure the leads and cables; keep the generator where it cannot be knocked.
  4. Have transcutaneous pads ready for a backup route.
  5. Have an experienced operator and an airway and drug plan.

If You Are Not the Device Specialist

  • Watch the ECG for capture and the pulse with each paced beat.
  • Do not disconnect the leads, change the mode or change the output.
  • If capture is lost or the patient is unstable, start the transcutaneous route per protocol and call the team.
  • Do not reposition a transvenous lead.

During Transport

  • Keep the patient on the monitor with the pacing settings recorded
  • Check the entry site and dressing at each handoff

Air Medical & Altitude

Battery life and device function at altitude depend on the manufacturer and on protocol. Check the transport team's device guidance before flight.

Handoff

  • Mode, rate and output; last capture check
  • Lead type and entry site; any repositioning
  • Anticoagulation status and recent platelet count
  • Recent alarms or complications
Transport Pearl: Pack the generator, the spare battery and the transcutaneous pads as one kit. Test the backup before you leave.

Special Populations

Pediatric & Neonatal

After pediatric cardiac surgery, epicardial wires are common. Pediatric thresholds, outputs and rates differ from adult values. They follow the pediatric cardiology protocol. The adult numbers on this page do not apply to children.

For the broader approach, see the pediatric shock page.

Pregnancy & Peripartum

Transvenous placement means fluoroscopy. Shield the abdomen and limit exposure per protocol. The team should talk through fetal dose with the radiology team. Pacing itself is managed the same way, with the obstetric team involved.

Bedside Pearls & Teaching Script

Systematic Walk-Through

  1. Indication: why is the patient bradycardic, and is pacing needed now?
  2. Route: transcutaneous, transvenous or epicardial?
  3. Mode and settings: as ordered.
  4. Capture: is there a QRS after each spike, and a pulse?
  5. Sensing: does the generator inhibit when the native rhythm returns?
  6. Complications and goal: what is the patient's plan today?

Pearls

  • The spike proves nothing until a QRS follows it.
  • A rising threshold is an early warning of a lead problem.
  • Never turn off the generator in a pacer-dependent patient.
  • The transcutaneous margin is a different number from the transvenous margin.

References

  1. Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Circulation. 2019 Aug 20. https://doi.org/10.1161/CIR.0000000000000628
  2. Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Heart Rhythm. 2019 Sep. https://doi.org/10.1016/j.hrthm.2018.10.037
  3. Epstein AE, DiMarco JP, Ellenbogen KA, et al. 2012 ACCF/AHA/HRS focused update of the 2008 guidelines for device-based therapy of cardiac rhythm abnormalities. Circulation. 2012. https://doi.org/10.1161/cir.0b013e3182618569
  4. Part 9: Adult Advanced Life Support: 2025 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16 suppl 2):S538–S577. https://doi.org/10.1161/cir.0000000000001376
  5. Chinese Medical Association Emergency Branch. Chinese emergency expert consensus on bedside temporary cardiac pacing (2023). Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2023. (In Chinese.) https://doi.org/10.3760/cma.j.cn121430-20230228-00124
  6. Atropine sulfate injection, prescribing information. DailyMed. Accessed 2026-10-15. https://dailymed.nlm.nih.gov/dailymed/drugInfo.cfm?setid=0f5b0961-efe7-42c3-898e-2f7fc7fc90b1
  7. Dopamine hydrochloride injection, prescribing information (Hospira). DailyMed. Accessed 2026-10-15. https://dailymed.nlm.nih.gov/dailymed/drugInfo.cfm?setid=1f306ad2-3606-4525-5a8a-ce78f426c1a2

Last updated Spotted something wrong? Report an error and include the page URL.

Medical Disclaimer
  • For Educational Purposes Only: This content is intended for educational reference and should not be used for clinical decision-making.
  • Not a Substitute for Professional Judgment: Always consult your local protocols, institutional guidelines, and supervising physicians.
  • Verify Before Acting: Users are responsible for verifying information through authoritative sources before any clinical application.