Bedside Topline
- What It Does
- A temporary pulse generator paces the heart through a transvenous catheter in the right ventricle or through epicardial wires placed at surgery.
- Key Settings & Targets
- Rate: 60–80/min (typical; follow the order)
- Output: above threshold by the protocol margin
- Capture: ECG and pulse confirmed
- Check Every Hour
- Capture on the ECG, pulse with each paced beat, rate and mode, battery, site and connections.
- Top Alarm → First Move
- Failure to capture: check the pulse, raise output in steps, check the connections, then call the team.
- Never
- Turn the generator off in a pacer-dependent patient; push or pull a lead; change the mode or output without an order.
- Escalate Now If
- Loss of capture with a falling blood pressure, pulseless electrical activity, muffled heart sounds, or a dislodged lead.
Overview & How It Works
An external pulse generator sends an electrical impulse to the heart. The route is a catheter placed in the right ventricle (transvenous) or wires sewn to the heart surface (epicardial). Use it when the heart cannot generate a rate or conduction that keeps the patient perfused. It bridges the time until a permanent solution or recovery. It is a bridge, not a fix.
Key Components
| Component | Where It Sits | What It Does |
|---|---|---|
| External pulse generator | Bedside, on a stand or in a transport case | Sets rate and output, senses the native rhythm, and shows the battery state |
| Transvenous pacing catheter | Right atrium and right ventricle, through an entry vein | Delivers the impulse at the right ventricular endocardium |
| Epicardial wires | Sewn to the atrial or ventricular epicardium at surgery, brought out through the chest wall | Deliver the impulse after cardiac surgery |
| Cables and connectors | Between the leads and the generator | Carry the signal; loose or wet connections cause sensing and capture faults |
| Entry site and dressing | Neck, chest, groin or the surgical wire exit site | Infection and dislodgement risk; check every shift |
Flow Path
- Generator: sets the rate, output and sensitivity.
- Lead: the impulse runs down the wire or catheter.
- Myocardium: once the impulse clears the threshold, the chamber depolarizes. That is capture.
- Sensing: it listens for the native beat, then inhibits or triggers as programmed.
Transcutaneous pacing runs on skin pads with its own settings. The transcutaneous pacing page covers that route.
Physiology & Hemodynamic Effects
What It Does
| Parameter | Effect | Why It Matters |
|---|---|---|
| Heart rate | ↑ to the set rate when the native rate is slower | Restores a rate that supports cardiac output in bradycardia or heart block |
| AV synchrony | ↑ with atrial-tracking modes (AAI, DDD) when the atria are intact | Atrial contraction adds to ventricular filling; ventricular-only pacing (VVI) loses it |
| Cardiac output | ↔ or ↑ when pacing restores a perfusing rate; ↓ with pacing-induced dyssynchrony | Output depends on rate, rhythm and filling, not on the spike alone |
| Ventricular activation | Altered (wide paced QRS) when pacing from the RV apex | Dyssynchrony can lower function in some patients, especially with reduced EF |
| Pacing threshold | ↑ early after insertion and with ischemia, acidosis, hyperkalemia or lead displacement | A rising threshold means the output may no longer capture |
What It Depends On
- Lead position and tissue contact. The threshold and sensing change when the lead moves.
- Electrolytes and acid-base status (potassium, pH, calcium)
- Native conduction. A pacer-dependent patient has no reliable escape rhythm to fall back on.
- Drugs that change the threshold or conduction (antiarrhythmics, some sedatives)
- Pacing mode and whether the atria are paced or only sensed
Types, Modes & Configurations
| Type / Mode | Route or Configuration | Best For | Key Limitation / Risk |
|---|---|---|---|
| Transvenous (RV) | Internal jugular, subclavian or femoral vein; tip in the RV apex | Bradycardia or heart block where the transcutaneous route fails or is not tolerated | Dislodgement, perforation, pneumothorax, infection, arrhythmia during placement |
| Epicardial (surgical) | Atrial and ventricular wires sewn at surgery; exit through the chest wall | Bradyarrhythmia after cardiac surgery, with the wires left in place | Wires are unipolar or bipolar by design; wrong connections cause failure to capture; removal can bleed |
| Transcutaneous (bridge) | Pads on the chest and back | Immediate bridge while a transvenous or epicardial route is arranged | Painful, high output, capture may be unreliable |
Pacing Mode Letters (NBG Code)
| Letter | Position | Meaning |
|---|---|---|
| 1st | Chamber paced | V = ventricle, A = atrium, D = both |
| 2nd | Chamber sensed | V, A or D; the generator looks for the native beat here |
| 3rd | Response to a sensed beat | I = inhibit (wait), T = trigger (pace now), D = both |
Common Modes
- VVI: ventricular pacing. A native ventricular beat inhibits it. This is the usual temporary mode.
- AAI: atrial pacing with intact AV conduction.
- DDD: dual-chamber. It tracks the atria while AV conduction is intact. It needs both leads.
- Asynchronous (VOO or DOO): it paces at a fixed rate and ignores the native rhythm. It is used to stop oversensing or inhibition during electrocautery. The trap: a spike that lands on a native T wave can cause R-on-T and VF.
Indications, Contraindications & Evidence
Indications
- Symptomatic bradycardia with poor perfusion that does not respond to drug therapy, per the local bradycardia algorithm
- Symptomatic Mobitz II or complete (third-degree) AV block, especially with a wide QRS or an unreliable escape rhythm
- Sinus node dysfunction with symptoms, as a bridge to a permanent device
- Bradyarrhythmia after cardiac surgery, with epicardial wires in place
- Bradycardia from drug toxicity, when drug therapy has failed and the team is arranging a pacing route
Contraindications
| Absolute | Relative (Protocol-Dependent) |
|---|---|
| No absolute contraindication in an emergency; the decision is to pace or not, based on the indication | Coagulopathy or thrombocytopenia (venous hematoma and bleeding) |
| Not applicable as a rule; some protocols list active line infection or bacteremia at the entry site | Tricuspid prosthetic valve, or other anatomy that makes placement difficult |
Pneumothorax risk at the subclavian or internal jugular site; lead placed by an experienced operator |
Limitations
- Transvenous pacing needs an operator who can place the lead and position it by fluoroscopy or ECG guidance
- Epicardial wires depend on where the surgeon placed them. Their function changes as the myocardium heals
- Temporary pacing does not treat the cause of the bradycardia
Key Evidence
| Guideline / Source (Year) | Population | Bottom Line |
|---|---|---|
| 2018 ACC/AHA/HRS bradycardia guideline (published 2019) | Adults with bradycardia and conduction delay | Covers the evaluation and management of symptomatic bradycardia and AV block, including temporary pacing, with its own recommendation classes |
| 2012 ACCF/AHA/HRS focused update of the 2008 device guideline | Adults needing cardiac rhythm devices | Updates device indications and pacing practice; read the recommendation tables before applying a class |
| 2025 AHA CPR and ECC guidelines, Part 9: Adult Advanced Life Support (2025) | Adults in cardiac arrest and peri-arrest conditions | Current AHA adult advanced life support guidance; confirm the bradycardia algorithm and drug doses in the full text before use |
| Chinese Medical Association Emergency Branch consensus on bedside temporary cardiac pacing (2023) | Emergency bedside temporary pacing | Covers route choice, venous access, lead selection, parameter settings, complications and post-procedure care (published in Chinese) |
Protocol-dependent. The margin for pacing output and the threshold to start pacing vary between institutions. This page sets no single number for either.
Setup, Settings & Targets
Targets
| Parameter | Typical Target | If Below | If Above |
|---|---|---|---|
| Rate | 60–80/min (typical; follow the order) | Check capture and sensing; call the team | Check for pacing-induced tachycardia or oversensing |
| Output (mA) | Set above the capture threshold by the protocol margin (protocol-dependent) | Failure to capture: raise output in steps and recheck the threshold | Output above what is needed increases the risk of tissue injury and generator drain |
| Sensitivity | Set so the generator sees the native beat and inhibits when it should | Undersensing: pacing competes with native beats | Oversensing: pacing is inhibited by noise or T waves |
| Pulse with each paced beat | Present when capture is confirmed | No pulse with spikes: pulseless electrical activity pathway | — |
Initial Settings
- Confirm the indication and the order; set the mode the order specifies.
- Set the rate; start at the ordered rate.
- Increase output in steps until capture is seen on the ECG and the pulse is checked.
- Decrease output in steps to find the threshold (the lowest output with capture); set the working output above the threshold by the protocol margin.
- Set sensitivity so the generator senses the native rhythm; test it by watching for inhibition with a native beat.
Sizing & Positioning
- Transvenous: fluoroscopy or the ECG pattern, plus a chest X-ray, confirm the tip position in the RV apex, per local protocol.
- Epicardial: the surgeon records which wires are atrial and which are ventricular. Keep that record with the generator.
Waveforms & Console Interpretation
Normal
- Capture looks like this: a spike, then a wide QRS, with the T wave pointing the opposite way, at the set rate.
- Demand pacing goes quiet when the native beat is there and sensed. No spike.
- In DDD, the atrial spike comes first. The ventricular spike follows after the programmed AV delay.
Abnormal Patterns
| Pattern | What You See | Cause | Fix |
|---|---|---|---|
| Failure to capture | Spike with no QRS after it | Output below threshold, lead displaced, perforation, hyperkalemia or acidosis, lead fracture | Check the pulse; raise output and recheck; check the lead and electrolytes; call the team |
| Failure to sense (undersensing) | Spikes fall on native beats; pacing competes | Sensitivity set too low, lead displaced | Adjust sensitivity; check the lead |
| Oversensing | Pauses; pacing inhibited by noise, T waves or muscle activity | Sensitivity set too high, electrical interference, loose connection | Adjust sensitivity; check the connections and cables |
| Spikes on the T wave | Spike falls in the vulnerable window of a native beat | Asynchronous mode or wrong settings | Turn to demand mode when the procedure ends; call the team |
| Battery or power alarm | Warning on the generator | Battery depletion, mains failure | Switch to backup power or a spare generator per protocol |
Monitoring & Ongoing Care
| Frequency | What to Check | Action Threshold |
|---|---|---|
| Hourly | Rhythm, capture on the ECG, pulse with paced beats, rate, generator mode, battery | Any loss of capture, pulse with spikes absent, or a rate change: check and call |
| Hourly | Insertion site, dressing, connections and cables, lead secured | Bleeding, drainage, loose connection, lead movement: call |
| Every 4–6 h | Threshold check per protocol; electrolytes (potassium, magnesium, calcium) | A rising threshold or abnormal electrolytes: treat per order and recheck |
| Daily | Chest X-ray or other imaging for lead position, per protocol | Position change: notify the team before any repositioning |
| Daily | Blood tests: platelets, aPTT or anti-Xa per order | Platelets or coagulation outside the order range: notify the team |
Positioning & Mobility
- Leads stay secured. On every move, protect them and keep the generator on the bed or in the transport case
- The entry site decides how you position the limb or neck. Avoid pulling on the cable
Anticoagulation & Hematology
| Strategy / Agent | Typical Target | Monitoring | Watch For |
|---|---|---|---|
| Unfractionated heparin (if ordered) | Per order and protocol | aPTT or anti-Xa per order | Venous hematoma, pericardial bleeding |
| Platelet count | Per protocol | Daily CBC | Thrombocytopenia increases hematoma risk |
| Antiplatelet therapy | Per order | Per order | Bleeding at the entry site |
Tag heparin High-alert. Refer to the heparin page and the aPTT page.
Bleeding & Platelets
- Venous puncture hematoma: pressure and a dressing, check the neck or groin each shift
- Pericardial bleeding: falling blood pressure and a rising heart rate after lead placement; call immediately
- Thrombocytopenia: check the platelet count (see the platelets page)
Key Medications & Drug Considerations
| Medication | Role Here | Dose / Target | Notes |
|---|---|---|---|
| Atropine | First drug for sinus and AV-nodal bradycardia before pacing, per the local bradycardia algorithm | Symptomatic bradycardia: give per the current AHA ACLS bradycardia algorithm and your local protocol, which set the dose, repeat interval and maximum. Label (coronary artery disease, section 2.4): limit the total dose to 0.03–0.04 mg/kg. Atropine is not recommended for asystole or PEA arrest. | Not a reliable step in Mobitz II or complete block; do not let it delay pacing |
| Dopamine | Infusion bridge while pacing is set up or when pacing fails | Bradycardia: infusion titrated to response per the current AHA ACLS bradycardia algorithm and your local protocol. The DailyMed label dose for shock (start 2–5 mcg/kg/min, maximum 50 mcg/kg/min) is for a different indication and is not a bradycardia dose. | High-alert infusion; use an infusion pump |
| Epinephrine infusion | Infusion bridge when dopamine is not used | Bradycardia: infusion titrated to response per the current AHA ACLS bradycardia algorithm and your local protocol | High-alert infusion |
| Heparin | Only if ordered for another indication | Per order | High-alert |
| Lidocaine | May be used for ventricular ectopy that the pacing lead provokes | Per order | Ventricular ectopy during placement is a reason to stop and reposition |
Alarms & Troubleshooting
| Alarm / Problem | What You See | Likely Causes | First Actions |
|---|---|---|---|
| Failure to capture | Spikes with no QRS | Output below threshold, lead displaced, electrolytes, acidosis, lead fracture or connection | 1) Check the pulse 2) Raise output in steps 3) Check connections and electrolytes 4) Call the team |
| Failure to sense | Spikes on top of native beats | Sensitivity too low, lead position | 1) Adjust sensitivity 2) Check the connections 3) Call the team |
| Oversensing | Pauses or inhibition with no native beat | Sensitivity too high, interference, loose cable | 1) Adjust sensitivity 2) Check the cables 3) Check for interference |
| Loss of capture after defibrillation | No capture after a shock | Threshold change from the shock, lead injury | 1) Check the pulse 2) Recheck the threshold 3) Call the team |
| Battery or power alarm | Generator warning | Depleted battery, power failure | 1) Switch to backup or spare generator 2) Keep pacing going 3) Call the team |
| Hypotension after pacing starts | Blood pressure falls after the mode changes | Wrong mode for the rhythm, ventricular-only pacing with good AV conduction | 1) Check the rhythm and mode 2) Call the team; mode changes need an order |
Complications & Red Flags
| Complication | How You Recognize It | Immediate Action | Prevention |
|---|---|---|---|
| Pneumothorax (neck or chest access) | Dyspnea, falling oxygen saturation, reduced breath sounds | Call the team; oxygen; chest X-ray | Experienced operator; imaging per protocol |
| Lead displacement | Failure to capture or sense, a changed ECG pattern | Check the output and capture; call before repositioning | Secure the lead; monitor the position |
| Perforation or cardiac tamponade | Hypotension, pulseless electrical activity, muffled heart sounds, pulsus paradoxus | Call the team now; prepare for pericardiocentesis per protocol | Careful placement with imaging; watch for threshold change |
| Infection | Fever, redness, drainage at the site | Notify the team; cultures per order | Sterile technique; dressing care per protocol |
| Venous thrombosis | Arm or neck swelling, pain | Call the team | Early removal when no longer needed |
| Arrhythmia during placement | Ventricular ectopy or VF during placement | Stop, call, reposition per the operator | Careful advancement; monitor the ECG |
Emergencies: Arrest, Power Loss & Device Failure
Cardiac Arrest on a Temporary Pacemaker
- Start CPR: follow the current ACLS algorithm. The pacemaker does not replace compressions.
- Pads: place defibrillation pads away from the pacing lead site per the manufacturer and protocol.
- Shock: defibrillate per the algorithm; some protocols disconnect the leads from the generator before a shock. Follow the local protocol.
- After the shock: check the pulse and the capture; recheck the threshold if pacing was lost.
Loss of Capture in an Unstable Patient
- Check the patient: pulse, blood pressure and mental status.
- Check the device: connections, output, battery.
- Bridge: start the transcutaneous route or the infusion bridge per order.
- Call: the team responsible for the lead.
Generator Failure in a Pacer-Dependent Patient
- Do not turn the generator off.
- Switch to a backup or spare generator per protocol.
- Bridge: transcutaneous pacing or the infusion per order.
- Call the team now.
Dislodgement or Suspected Perforation
- Do not push or pull the lead.
- Cover the entry site with a sterile dressing.
- Call the team; watch for hypotension and tamponade signs.
Weaning & Discontinuation
Readiness Criteria
- The native rhythm is stable with an adequate escape rate, per the order
- Capture and sensing are stable on the current settings
- The cause of the bradycardia has resolved or is controlled
- The electrolytes are corrected
Weaning Steps
- Lower the rate in steps, or switch to demand mode, per the order.
- Watch the native rhythm and perfusion at each step.
- Pause pacing only under an order, with a team present and a backup ready.
Failure Criteria: Step Back Up
- Falling blood pressure, altered mental status or chest pain with the native rate
- Pauses or junctional or idioventricular escape that is too slow
Removal & Post-Removal Care
- Removal per the order and the local protocol
- The surgical team removes epicardial wires per protocol. Do not pull against resistance
- Watch the site for bleeding. Check the chest X-ray per protocol after a transvenous removal
Transport Considerations
Before You Move
- Check battery charge and carry a charged spare.
- Confirm the mode, rate and output with the sending team; write them down.
- Secure the leads and cables; keep the generator where it cannot be knocked.
- Have transcutaneous pads ready for a backup route.
- Have an experienced operator and an airway and drug plan.
If You Are Not the Device Specialist
- Watch the ECG for capture and the pulse with each paced beat.
- Do not disconnect the leads, change the mode or change the output.
- If capture is lost or the patient is unstable, start the transcutaneous route per protocol and call the team.
- Do not reposition a transvenous lead.
During Transport
- Keep the patient on the monitor with the pacing settings recorded
- Check the entry site and dressing at each handoff
Air Medical & Altitude
Battery life and device function at altitude depend on the manufacturer and on protocol. Check the transport team's device guidance before flight.
Handoff
- Mode, rate and output; last capture check
- Lead type and entry site; any repositioning
- Anticoagulation status and recent platelet count
- Recent alarms or complications
Special Populations
Pediatric & Neonatal
After pediatric cardiac surgery, epicardial wires are common. Pediatric thresholds, outputs and rates differ from adult values. They follow the pediatric cardiology protocol. The adult numbers on this page do not apply to children.
For the broader approach, see the pediatric shock page.
Pregnancy & Peripartum
Transvenous placement means fluoroscopy. Shield the abdomen and limit exposure per protocol. The team should talk through fetal dose with the radiology team. Pacing itself is managed the same way, with the obstetric team involved.
Bedside Pearls & Teaching Script
Systematic Walk-Through
- Indication: why is the patient bradycardic, and is pacing needed now?
- Route: transcutaneous, transvenous or epicardial?
- Mode and settings: as ordered.
- Capture: is there a QRS after each spike, and a pulse?
- Sensing: does the generator inhibit when the native rhythm returns?
- Complications and goal: what is the patient's plan today?
Pearls
- The spike proves nothing until a QRS follows it.
- A rising threshold is an early warning of a lead problem.
- Never turn off the generator in a pacer-dependent patient.
- The transcutaneous margin is a different number from the transvenous margin.
References
- Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Circulation. 2019 Aug 20. https://doi.org/10.1161/CIR.0000000000000628
- Kusumoto FM, Schoenfeld MH, Barrett C, et al. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay. Heart Rhythm. 2019 Sep. https://doi.org/10.1016/j.hrthm.2018.10.037
- Epstein AE, DiMarco JP, Ellenbogen KA, et al. 2012 ACCF/AHA/HRS focused update of the 2008 guidelines for device-based therapy of cardiac rhythm abnormalities. Circulation. 2012. https://doi.org/10.1161/cir.0b013e3182618569
- Part 9: Adult Advanced Life Support: 2025 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2025;152(16 suppl 2):S538–S577. https://doi.org/10.1161/cir.0000000000001376
- Chinese Medical Association Emergency Branch. Chinese emergency expert consensus on bedside temporary cardiac pacing (2023). Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2023. (In Chinese.) https://doi.org/10.3760/cma.j.cn121430-20230228-00124
- Atropine sulfate injection, prescribing information. DailyMed. Accessed 2026-10-15. https://dailymed.nlm.nih.gov/dailymed/drugInfo.cfm?setid=0f5b0961-efe7-42c3-898e-2f7fc7fc90b1
- Dopamine hydrochloride injection, prescribing information (Hospira). DailyMed. Accessed 2026-10-15. https://dailymed.nlm.nih.gov/dailymed/drugInfo.cfm?setid=1f306ad2-3606-4525-5a8a-ce78f426c1a2
Medical Disclaimer
- For Educational Purposes Only: This content is intended for educational reference and should not be used for clinical decision-making.
- Not a Substitute for Professional Judgment: Always consult your local protocols, institutional guidelines, and supervising physicians.
- Verify Before Acting: Users are responsible for verifying information through authoritative sources before any clinical application.